冷凍離(li)(li)心(xin)機相(xiang)較于一般(ban)的離(li)(li)心(xin)機,顯著作用就(jiu)是在離(li)(li)心(xin)過程中對溫度進行控(kong)制,以(yi)保證樣品的生物活性。如(ru)果冷凍(dong)離心機制冷出(chu)現故障或者效(xiao)果(guo)差,就會導致實(shi)驗(yan)達不到(dao)預(yu)期效(xiao)果(guo)。我(wo)們來分析一下,在日常使(shi)用中(zhong),導致冷凍離心機制冷效(xiao)果(guo)差的原因(yin)有哪些。
1. 離心機通風性能不好,如儀器安(an)裝在日光直射或通風不良處(chu),散熱(re)器效(xiao)果差,或散熱(re)器蓋滿灰塵,會影響制冷效(xiao)果。
2. 電(dian)源(yuan)不通是導致冷(leng)凍離心機(ji)啟(qi)動及制冷(leng)效果差的原(yuan)因之一,與電(dian)機(ji)不轉的故障相同,應(ying)分別檢查(cha)電(dian)源(yuan)及保險(xian)絲。
3. 離心機(ji)離心管重量(liang)不(bu)平(ping)(ping)衡,放置(zhi)不(bu)對稱; 轉子孔內有(you)異物,使負(fu)荷(he)不(bu)平(ping)(ping)衡; 轉軸(zhou)上端(duan)固(gu)定(ding)螺(luo)帽(mao)松動,轉軸(zhou)磨擦或彎曲; 電機(ji)轉子不(bu)在磁場中心會(hui)產(chan)生噪音(yin); 轉子本(ben)身損傷等都可導致機(ji)體震動劇烈、響聲異常(chang)。
4. 電(dian)(dian)(dian)壓過低(di)(di)(di),安全裝置故障(zhang)也可使冷凍離(li)心機不(bu)(bu)能啟(qi)動,電(dian)(dian)(dian)壓過低(di)(di)(di)可能是電(dian)(dian)(dian)網電(dian)(dian)(dian)壓低(di)(di)(di); 配(pei)電(dian)(dian)(dian)板配(pei)線過多(duo); 電(dian)(dian)(dian)源(yuan)線過長等(deng)。電(dian)(dian)(dian)源(yuan)線容量不(bu)(bu)夠,不(bu)(bu)僅(jin)使電(dian)(dian)(dian)壓降低(di)(di)(di),還會造(zao)成意外(wai)事故。當電(dian)(dian)(dian)源(yuan)電(dian)(dian)(dian)壓降到180-190VJ 時,冷凍離(li)心機不(bu)(bu)能啟(qi)動,影(ying)響制冷效果。
5、環境溫度(du)(du)過高,離心室內溫度(du)(du)過高。
6、假如修理過的設備,那么可能是修理后重新灌入的冷媒量過多或過少或冷媒不符合要求。